Agilent Technologies Signal Analyzer N9030a User Manual
Page 1214
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
Remote Command
:TRIGger[:SEQuence]:RFBurst:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:RFBurst:SLOPe?
Example
TRIG:RFB:SLOP NEG
Couplings
This same slope is used in the RF Burst selection for the trigger source in the Trigger menu and
for the period timer sync source (in the Trigger menu and in the Gate Source menu).
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:RFBurst:SLOPe
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.
Initial S/W Revision
Prior to A.02.00
Periodic Timer (Frame Trigger)
Pressing this key, when it is not selected, selects the internal periodic timer signal as the trigger.
Triggering occurrences are set by the Period parameter, which is modified by the Sync Source
and Offset. Pressing this key, when it is already selected, accesses the periodic timer trigger
setup functions.
If you do not have a sync source selected (it is Off), then the internal timer will not be
synchronized with any external timing events.
Key Path
Trigger
Example
TRIG:SOUR FRAM Swept SA measurement
TRIG:
State Saved
Saved in instrument state
Readback
[Sync:
Status Bits/OPC
dependencies
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.
Initial S/W Revision
Prior to A.02.00
Periodic Timer Triggering:
1214
Remote Language Compatibility Measurement Application Reference