Agilent Technologies Signal Analyzer N9030a User Manual
Page 1216
![background image](/manuals/40104/1216/background.png)
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
Period
Sets the period of the internal periodic timer clock. For digital communications signals, this is
usually set to the frame period of your current input signal. In the case that sync source is not set
to OFF, and the external sync source rate is changed for some reason, the periodic timer is
synchronized at the every external synchronization pulse by resetting the internal state of the
timer circuit.
Key Path
Trigger, Periodic Timer
Remote Command
:TRIGger[:SEQuence]:FRAMe:PERiod
:TRIGger[:SEQuence]:FRAMe:PERiod?
Example
TRIG:FRAM:PER 100 ms
Dependencies
The invalid data indicator turns on when the period is changed, until the next
sweep/measurement completes.
Couplings
The same period is used in the Gate Source selection of the period timer.
Preset
20 ms
GSM: 4.615383
State Saved
Saved in instrument state
Min
100.000 ns
Max
559.0000 ms
Default Unit
S
Initial S/W Revision
Prior to A.02.00
1216
Remote Language Compatibility Measurement Application Reference