6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control

Period

Sets the period of the internal periodic timer clock. For digital communications signals, this is
usually set to the frame period of your current input signal. In the case that sync source is not set
to OFF, and the external sync source rate is changed for some reason, the periodic timer is
synchronized at the every external synchronization pulse by resetting the internal state of the
timer circuit.

Key Path

Trigger, Periodic Timer

Remote Command

:TRIGger[:SEQuence]:FRAMe:PERiod

:TRIGger[:SEQuence]:FRAMe:PERiod?

Example

TRIG:FRAM:PER 100 ms

Dependencies

The invalid data indicator turns on when the period is changed, until the next
sweep/measurement completes.

Couplings

The same period is used in the Gate Source selection of the period timer.

Preset

20 ms

GSM: 4.615383

State Saved

Saved in instrument state

Min

100.000 ns

Max

559.0000 ms

Default Unit

S

Initial S/W Revision

Prior to A.02.00

1216