Agilent Technologies Signal Analyzer N9030a User Manual
Page 1265
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
System
Alert
The instrument will signal an Alert when conditions exist such that you will need to perform a
full alignment (for example, Align Now, All). The Alert can be configured in one of four settings;
Time & Temperature, 24 hours, 7 days, or None. A confirmation is required when a selection other
than Time & Temperature is chosen. This prevents accidental deactivation of alerts.
With Auto Align set to Normal, the configuration of Alert is not relevant because the instrument’s
software maintains the instrument in warranted operation.
Key Path
System, Alignments, Auto Align
Mode
All
Remote Command
:CALibration:AUTO:ALERt TTEMperature|DAY|WEEK|NONE
:CALibration:AUTO:ALERt?
Example
:CAL:AUTO:ALER TTEM
Notes
The alert that alignment is needed is the setting of bit 14 in the Status Questionable Calibration
register.
Preset
This is unaffected by Preset but is set to TTEMperature on a “Restore System Defaults->Align”.
State Saved
No
Status Bits/OPC
dependencies
The alert is the Error Condition message “Align Now, All required” and bit 14 is set in the Status
Questionable Calibration register.
Initial S/W Revision
Prior to A.02.00
Time & Temperature
With Auto Align Alert set to Time & Temperature the instrument will signal an alert when
alignments expire due to the combination of the passage of time and changes in temperature. The
alert is the Error Condition message “Align Now, All required”. If this choice for Alert is selected,
the absence of an alert means that the analyzer alignment is sufficiently up-to-date to maintain
warranted accuracy.
Key Path
System, Alignments, Auto Align, Alert
Mode
All
Example
:CAL:AUTO:ALER TTEM
Readback Text
Time & Temp
Status Bits/OPC
dependencies
Bit 14 is set in the Status Questionable Calibration register.
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference
1265