6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
parameter is honored and the HIGH state set, and returned to a query, but the Medium dither
level is actually used.
Preset
AUTO
Backwards
Compatibility SCPI
The old command
[:SENSe]:ADC:DITHer AUTO
is aliased to
[:SENSe]:ADC:DITHer:AUTO[:STATe] ON; because of this, the [:SENSe]:ADC:DITHer function
cannot be a true Boolean, so the query, [:SENSe]:ADC:DITHer? returns OFF or ON (not 1 or 0 like
a true Boolean)
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.02.00
Auto
Sets the ADC dither to automatic. The analyzer then chooses the dither level according to which is
most likely to be the best selection, based on other settings within the digital IF.
When in Auto, the analyzer sets the dither to Medium whenever the effective IF Gain is Low by
this definition of IF Gain = Low:
•
When Sweep Type = Swept, IF Gain = Low whenever Swept IF Gain is set to Low Gain, whether
by autocoupling or manual selection.
•
When Sweep Type = FFT, IF Gain = Low whenever FFT IF Gain is set to "Low Gain," which
cannot happen by autocoupling.
Whenever the IF Gain is not low by this definition, Auto sets the dither to Off.
Key Path
Meas Setup, ADC Dither
Remote Command
[:SENSe]:ADC:DITHer:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:ADC:DITHer:AUTO[:STATe]?
Example
ADC:DITH:AUTO ON
Preset
ON
State Saved
Saved in instrument state
Readback
The “Auto” is underlined, and the readback value is whatever setting is auto-selected
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference
839