6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
Preset
AUTO
Backwards
Compatibility SCPI
[:SENSe]:SWEep:TYPE AUTO
sets sweep type Auto to On but the query will return either FFT or SWE depending on the auto
setting.
[:SENSe]:SWEep:TYPE SWP
selects sweep type Swept but will return SWE on a query
Initial S/W Revision
Prior to A.02.00
Auto
When in Auto, the selection of sweep type is governed by two different sets of rules, depending on
whether you want to optimize for dynamic range or for speed. These rules are chosen under the
Sweep Type Rules
key.
Key Path
Sweep/Control, Sweep Setup, Sweep Type
Remote Command
[:SENSe]:SWEep:TYPE:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TYPE:AUTO?
Example
:SWE:TYPE:AUTO ON
Couplings
Pressing Auto Couple always sets Sweep Type to Auto.
Swept is always chosen whenever any form of Signal ID is on, or the Source Mode is set to
Tracking, or any EMI detector is selected, or the RF Preselector is ON.
Preset
ON
State Saved
Saved in instrument state
Initial S/W Revision
Prior to A.02.00
Swept
Manually selects swept analysis, so it cannot change automatically to FFT.
Key Path
Sweep/Control, Sweep Setup, Sweep Type
Example
SWE:TYPE SWE
Dependencies
Grayed out while in Gated FFT (meaning Gate is ON and Gate Method is FFT).
If this key is selected, the gate method Gated FFT is grayed out.
Couplings
This selection is chosen automatically if any of the CISPR detectors is chosen for any active
trace, in which case the FFT Sweep Type selection is also grayed out.