Agilent Technologies Signal Analyzer N9030a User Manual
Page 1411
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Compatibility SCPI
Initial S/W Revision
Prior to A.02.00
Trig Slope
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.
Key Path
Trigger, External 2
Remote Command
:TRIGger[:SEQuence]:EXTernal2:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:EXTernal2:SLOPe?
Example
TRIG:EXT2:SLOP NEG
Couplings
This same slope is used in the Ext2 selection for the trigger source in the Trigger menu and for
the period timer sync source (in the Trigger menu and in the Gate Source menu).
Preset
POSitive
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal2:SLOPe
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the VID, LINE, EXT1,
EXT2, and RFB triggers.
Initial S/W Revision
Prior to A.02.00
Trig Delay
Controls a time delay during which the analyzer will wait to begin a sweep after meeting the
trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT,
but not in swept spans.
Key Path
Trigger, External 2
Remote Command
:TRIGger[:SEQuence]:EXTernal2:DELay
:TRIGger[:SEQuence]:EXTernal2:DELay?
:TRIGger[:SEQuence]:EXTernal2:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:EXTernal2:DELay:STATe?
Example
TRIG:EXT2:DEL:STAT ON
TRIG:EXT2:DEL 100 ms
Remote Language Compatibility Measurement Application Reference
1411