Ca (couple attenuation), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 254: Description, Syntax legacy products
5 Legacy Command Descriptions
CA (Couple Attenuation)
CA (Couple Attenuation)
Syntax
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
During normal operation, the instrument’s input attenuation is coupled to the reference level. This
coupling keeps the mixer input at a level such that a continuous wave signal displayed at the
reference level is at or below –10 dBm (or the value specified in the ML command.)
The CA command sets the threshold to –10 dBm (or to the value specified by the commands
or
"KS, (Mixer Level)" on page 318
). The counterpart to the CA
command is
"AT (Input Attenuation)" on page 236
, which allows levels less than the threshold
value at the mixer input.
Format
CA
Query Data Type
N/A
SCPI Equivalent
Commands
[:SENse]:POWer[:RF]:ATTenuation:AUTO ON (see
)
254
Remote Language Compatibility Measurement Application Reference