Agilent Technologies Signal Analyzer N9030a User Manual
Page 153
3 Programming the Analyzer
STATus Subsystem
Questionable Calibration Extended Failure Event Query
This query returns the decimal value of the sum of the bits in the Questionable Calibration
Extended Failure Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:EXTended:FAILure[:EVENt]?
Example
STAT:QUES:CAL:EXT:FAIL?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Extended Failure Negative Transition
This command determines which bits in the Questionable Calibration Extended Failure Condition
register will set the corresponding bit in the Questionable Calibration Extended Failure Event
register when the condition register bit has a negative transition (1 to 0). The variable
is the sum of the decimal values of the bits that you want to enable.
Mode
All
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:FAILure:NTRansition
:STATus:QUEStionable:CALibration:EXTended:FAILure:NTRansition?
Example
STAT:QUES:CAL:EXT:FAIL:NTR 1
EMI conducted align failure is not required.
Preset
0
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference
153