Agilent Technologies Signal Analyzer N9030a User Manual
Page 145
3 Programming the Analyzer
STATus Subsystem
dependencies
Initial S/W Revision
Prior to A.02.00
Questionable Event Query
This query returns the decimal value of the sum of the bits in the Questionable Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode
All
Remote Command
:STATus:QUEStionable[:EVENt]?
Example
STAT:QUES?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Negative Transition
This command determines which bits in the Questionable Condition register will set the
corresponding bit in the Questionable Event register when the condition register bit has a
negative transition (1 to 0). The variable
that you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:NTRansition 16
:STATus:QUEStionable:NTRansition?
Example
STAT:QUES:NTR 16 Temperature summary ‘questionable cleared’ will be reported to the Status
Byte Register.
Preset
0
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference
145