6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
at the bottom of the bandwidth will “fold”. For example, with a 40 MHz bandwidth (20 MHz half-
bandwidth), and a 15 MHz IF center, a signal –20 MHz relative to the spectrum analyzer center
frequency will have a relative response of about –3 dB with a frequency 20 MHz below the 15 MHz
IF center. This –5 MHz frequency will fold to become a +5 MHz signal at the IF output. Therefore,
lower IF output frequencies are only useful with known band-limited signals.
Key Path
Input/Output, Output Config, Aux IF Out
Example
OUTP:AUX AIF
causes the aux output type to be the Arbitrary IF
Dependencies
Does not appear unless Option CRP is installed.
Readback Text
Arbitrary IF
Initial S/W Revision
A.04.00
Key Path
Input/Output, Output Config, Aux IF Out
Scope
Mode Global
Remote Command
:OUTPut:AUX:AIF
:OUTPut:AUX:AIF?
Example
:OUTP:AUX:AIF 50 MHZ
Preset
This is unaffected by a Preset but is set to 70 MHz on a "Restore Input/Output Defaults" or
"Restore System Defaults->All”
State Saved
Saved in Input/Output State
Min
10 MHz
Max
75 MHz
Default Unit
Hz
Initial S/W Revision
A.04.00
Fast Log Video
In this mode the 2nd IF output is passed through a log amp and the log envelope of the IF signal
is sent to the rear panel. The open circuit output level varies by about 25 mV per dB, with a top-
of-screen signal producing about 1.6 Volts. The output impedance is nominally 50 ohms.
This mode is intended to meet the same needs as Option E4440A-H7L Fast Rise Time Video
Output on the Agilent E4440A PSA Series, allowing you to characterize pulses with fast rise times
using standard measurement suites on modern digital scopes.
Remote Language Compatibility Measurement Application Reference