External 1 – Agilent Technologies Signal Analyzer N9030a User Manual
Page 1407
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Default Unit
S
Backwards Compatibility
Notes
The legacy :TRIGger[:SEQuence]:DELay command affects the delay for the VID, LINE, EXT1,
EXT2, and RFB triggers.
The legacy :TRIGger[:SEQuence]:OFFSet command is supported for the VIDeo, LINE, EXT1, and
EXT2 triggers.
Initial S/W Revision
Prior to A.02.00
External 1
Pressing this key, when it is not selected, selects an external input signal as the trigger. A new
sweep/measurement will start when the external trigger condition is met using the external 1
input connector on the rear panel.
Pressing this key, when it is already selected, accesses the external 1 trigger setup menu.
Key Path
Trigger
Example
TRIG:SOUR EXT1 Swept SA measurement
TRIG:
Dependencies
Grayed out if in use by Point Trigger in the Source Setup menu.
Forced to Free Run if already selected and Point Trigger is set to External 1.
State Saved
Saved in instrument state
Status Bits/OPC
dependencies
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.
Initial S/W Revision
Prior to A.02.00
Trigger Level
Sets the value where the external 1 trigger input will trigger a new sweep/measurement.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:LEVel
:TRIGger[:SEQuence]:EXTernal1:LEVel?
Remote Language Compatibility Measurement Application Reference
1407