6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
Example
SENS:CORR:CSET1 ON
Dependencies
Turning this on automatically turns on "Apply Corrections"
Only the first correction array (Correction 1) supports antenna units. When this array is turned
on, and it contains an Antenna Unit other than “None”, the Y Axis Unit of the analyzer is forced
to that Antenna Unit. All other Y Axis Unit choices are grayed out.
Note that this means that a correction file with an Antenna Unit can only be loaded into the
Corrections 1 register. Consequently only for Correction 1 does the dropdown in the Recall dialog
include.ant, and if an attempt is made to load a correction file into any other Correction register
which DOES contain an antenna unit, a Mass Storage error is generated.
This command will generate an “Option not available” error unless you have the proper option
installed in your instrument.
Preset
Not affected by a Preset. Set to OFF by Restore Input/Output Defaults
State Saved
Saved in instrument state.
Backwards Compatibility
Notes
Unlike legacy analyzers, Preset does not turn Corrections off (Restore Input/Output Defaults
does).
Initial S/W Revision
A.02.00
Properties
Accesses a menu that lets you set the properties of the selected correction.
Key Path
Input/Output, Corrections
Initial S/W Revision
A.02.00
Select Correction
Specifies the selected correction. The term "selected correction" is used throughout this document
to specify which correction will be affected by the functions.
Key Path
Input/Output, Corrections
Mode
SA
Notes
The selected correction is remembered even when not in the correction menu.
Preset
Set to Correction 1 by Restore Input/Output Defaults
Readback
Correction 1|Correction 2|Correction 3|Correction 4|Correction 5|Correction 6
Initial S/W Revision
A.02.00