6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker Function
Preset
Quasi Peak
State Saved
Saved in instrument state
Readback Text
Detector name
Backwards
Compatibility SCPI
[:SENSe]:EMI:MEASure:DETector:QPEak[:STATe] OFF|ON|0|1|
This command is included for compatibility with the E7400 and PSA option 239 . If sent with On
as a parameter, sets detector 2 to Quasi Peak. If sent with Off as a parameter, sets detector 2 to
Off
Initial S/W Revision
A.02.00
Detector 2
This menu lets you select the detector to be used for Detector 2, or turn Detector 2 off. This is a
1-of-N menu that shows the normal list of detectors, but with the “Auto” key replaced by “Off”.
Key Path
Marker Function, Measure at Marker, Detectors
Remote Command
See
Example
:CALC:MAM:DET2 QPE
Sets the detector for measure at marker detector 2 to Quasi peak
:CALC:MAM:DET2 OFF
Sets the detector for measure at marker detector 2 to Off
Preset
Quasi Peak
State Saved
Saved in instrument state
Readback Text
Detector name
Backwards
Compatibility SCPI
[:SENSe]:EMI:MEASure:DETector:QPEak[:STATe] OFF|ON|0|1|
This command is included for compatibility with the E7400 and PSA option 239 . If sent with On
as a parameter, sets detector 2 to Quasi Peak. If sent with Off as a parameter, sets detector 2 to
Off
Initial S/W Revision
A.02.00
Detector 3
This menu lets you select the detector to be used for Detector 3, or turn Detector 3 off. This is a
1-of-N menu that shows the normal list of detectors, but with the “Auto” key replaced by “Off”.
Key Path
Marker Function, Measure at Marker, Detectors
Remote Command
See