Fpka (fast preselector peak), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 298: Description, Syntax legacy products
5 Legacy Command Descriptions
FPKA (Fast Preselector Peak)
FPKA (Fast Preselector Peak)
Syntax
Legacy Products
8566A/B
Description
Automatically adjusts the preselector frequency to yield the greatest signal level at the active
marker. The FPKA command peaks the preselector faster than the preselector-peak command, PP
Although this command can be executed in all frequency spans, it performs best when the
instrument is in zero span. Use the standard preselector peak for all other frequency spans.
The FPKA command also returns the amplitude value of active marker.
Format
FPKA
Query Data Type
Amplitude value of active marker.
SCPI Equivalent Commands
[:SENSe]:POWer[:RF]:PCENter (see
298
Remote Language Compatibility Measurement Application Reference