6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
Example
SWE:EGAT:HOLD 0.0002
SWE:EGAT:HOLD?
SWE:EGAT:HOLD:AUTO ON
SWE:EGAT:HOLD:AUTO?
Couplings
When Gate Holdoff is Auto, the Gate Holdoff key shows the value calculated by the analyzer for
the wait time.
Pressing the Gate Holdoff key while it is in Auto and not selected, causes the key to become
selected and allows the user to adjust the value. If the value is adjusted, the setting changes to
Man.
Pressing the Gate Holdoff key, while it is in Auto and selected, does not change the value of
Gate Holdoff, but causes the setting to change to Man. Now the user can adjust the value.
Pressing the key while it is in Man and selected, cause the value to change back to Auto.
Pressing the key while it is in Man and not selected, causes the key to become selected and
allows the user to adjust the value.
When Method is set to Video or FFT, the Gate Holdoff function has no effect.
Preset
Auto
Auto/On
State Saved
Saved in instrument state
Min
1 µsec
Max
1 sec
Initial S/W Revision
Prior to A.02.00
Min Fast Position Query (Remote Command Only)
This command queries the position of the MIN FAST line, relative to the delay reference (REF)
line. See section
"Gate View On/Off" on page 1197
. If this query is sent while not in gate view, the
MinFast calculation is performed based on the current values of the appropriate parameters and
the result is returned. Knowing this value lets you set an optimal gate delay value for the current
measurement setup.
Remote Command
[:SENSe]:SWEep:EGATe:MINFast?
Example
SWE:EGAT:MIN?
Initial S/W Revision
Prior to A.02.00
Gate Preset (Remote Command Only)
Presets the time-gated spectrum analysis capability.
Remote Language Compatibility Measurement Application Reference