Agilent Technologies Signal Analyzer N9030a User Manual
Page 620
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
Key Path
Input/Output, External Gain
Remote Command
[:SENSe]:CORRection:SA[:RF]:GAIN
[:SENSe]:CORRection:SA[:RF]:GAIN?
Example
CORR:SA:GAIN 10 sets the Ext Gain value to 10 dB
CORR:SA:GAIN –10 sets the Ext Gain value to –10 dB (that is, an attenuation of 10 dB)
Notes
Does not auto return.
Dependencies
The reference level limits are determined in part by the External Gain/Atten, Max Mixer Level,
and RF Atten.
This key is grayed out in Modes that do not support External Gain
Preset
This is unaffected by Preset but is set to 0 dB on a "Restore Input/Output Defaults" or "Restore
System Defaults->All"
State Saved
Saved in instrument state
Min
–120 dB
Max
120 dB
Readback
Preamp Gain,
Backwards
Compatibility SCPI
[:SENSe]:CORRection:OFFSet[:MAGNitude]
The legacy "Ext Preamp Gain" key is now called "Ext Gain" and the sub-menu has choices of
Ext Preamp | MS | BTS for backwards compatibility.
Initial S/W Revision
Prior to A.02.00
MS
Sets an external gain/attenuation value for MS (Mobile Station) tests.
Key Path
Input/Output, External Gain
Remote Command
[:SENSe]:CORRection:MS[:RF]:GAIN
[:SENSe]:CORRection:MS[:RF]:GAIN?
Example
CORR:MS:GAIN 10 sets the Ext Gain value to 10 dB
CORR:MS:GAIN –10 sets the Ext Gain value to –10 dB (that is, a loss of 10 dB.)
Notes
Does not auto return.
Dependencies
The reference level limits are determined in part by the External Gain, Max Mixer Level, RF Atten
This key is grayed out in modes that do not support MS.
Preset
This is unaffected by a Preset but is set to 0 dB on a "Restore Input/Output Defaults" or
"Restore System Defaults->All"
620
Remote Language Compatibility Measurement Application Reference