6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker Function
If the Measure at Marker window is being displayed, and one of the detectors is changed, any
value being displayed for that detector changes to “---“ until the next successful reading from that
detector.
Key Path
Marker Function, Measure at Marker,
Remote Command
:CALCulate:MAMarker:DETector[1] | 2 | 3 OFF | NORMal | AVERage |
POSitive | SAMPle | NEGative | QPEak | EAVerage | RAVerage
:CALCulate:MAMarker:DETector[1]|2|3?
Example
:
CALC:MAM:DET2 QPE
Sets the detector for measure at marker detector 2 to Quasi peak
:CALC:MAM:DET OFF
Sets the detector for measure at marker detector 1 to Off
State Saved
Saved in instrument state
Initial S/W Revision
A.02.00
Key Path
Marker Function, Measure at Marker,
Remote Command
:CALCulate:MAMarker:DETector[1]|2|3:DWELl
:CALCulate:MAMarker:DETector[1]|2|3:DWELl?
Example
:CALC:MAM:DET2:DWEL 500 ms
Sets the detector for measure at marker detector 2 to dwell for 500 ms
State Saved
Saved in instrument state
Backwards
Compatibility SCPI
[:SENSe]:EMI:MEASure:DETector:DWELl
This command is included for compatibility with the E7400 and PSA option 239 . Sets all of the
detectors’ dwell times to the specified amount
Initial S/W Revision
A.02.00
Detector 1
This menu lets you select the detector to be used for Detector 1, or turn Detector 1 off. This is a
1-of-N menu that shows the normal list of detectors, but with the “Auto” key replaced by “Off”.
Key Path
Marker Function, Measure at Marker, Detectors
Remote Command
See
Example
:CALC:MAM:DET QPE
Sets the detector for measure at marker detector 1 to Quasi peak
:CALC:MAM:DET OFF