Ks, (mixer level), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 318: Description, Syntax legacy products
5 Legacy Command Descriptions
KS, (Mixer Level)
KS, (Mixer Level)
Syntax
Legacy Products
8566A/B, 8568A/B
Description
Specifies the maximum signal level that is applied to the input mixer for a signal that is equal to
or below the reference level.
The effective mixer level is equal to the reference level minus the input attenuator setting. When
KS, is activated, the effective mixer level can be set from –10 dBm to –70 dBm in 10 dB steps.
As the reference level is changed, the coupled input attenuator automatically changes to limit the
maximum signal at the mixer input to your specified setting for signals less than or equal to the
reference level.
Format
KS,
KS, OA
KS,?
Query Data Type
SCPI Equivalent Commands
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]
:[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
(See
Preset
– 10 dBm
Notes
The functions of KS, are identical to
"ML (Mixer Level)" on page 411
.
If the external amplifier gain has been set, executing IP does not reset this value. This is
to protect the instrument.
318
Remote Language Compatibility Measurement Application Reference