Agilent Technologies Signal Analyzer N9030a User Manual
Page 156
3 Programming the Analyzer
STATus Subsystem
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Extended Needed Negative Transition
This command determines which bits in the Questionable Calibration Extended Needed Condition
register will set the corresponding bit in the Questionable Calibration Extended Needed Event
register when the condition register bit has a negative transition (1 to 0). The variable
is the sum of the decimal values of the bits that you want to enable.
Mode
All
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition
:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition?
Example
STAT:QUES:CAL:EXT:NEED:NTR 2
Align EMI conducted is not required.
Preset
0
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Questionable Calibration Extended Needed Positive Transition
This command determines which bits in the Questionable Calibration Extended Needed Condition
register will set the corresponding bit in the Questionable Calibration Extended Needed Event
register when the condition register bit has a positive transition (0 to 1). The variable
the sum of the decimal values of the bits that you want to enable.
Mode
All
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:PTRansition
:STATus:QUEStionable:CALibration:EXTended:NEEDed:PTRansition?
Example
STAT:QUES:CAL:EXT:NEED:PTR 2
Align EMI conducted is required.
156
Remote Language Compatibility Measurement Application Reference