Mkrδ->cf, Mkrδ->span, Mkr ->cf – Agilent Technologies Signal Analyzer N9030a User Manual
Page 782: Mkr ->span
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker To
Remote Command
:CALCulate:MARKer[1]|2|...12[:SET]:ZSPan:CENTer
Example
:CALC:MARK2:ZSP:CENT sets the Zone CF to the value of marker 2.
Notes
Sending this command selects the subopcoded marker
If specified marker is off, this command will first turn it on at the center of the screen as a
normal type marker. Then the Mkr->Zone Center function is performed.
Dependencies
Only appears in the Zone Span View of the Swept SA measurement. If the SCPI command is
sent in other Views, gives an error.
In addition, this function is not available when the bottom window is in Zero Span.
Initial S/W Revision
A.07.01
Mkr ->CF
Sets the center frequency to the frequency difference between the selected marker and its reference
marker. The marker is then changed to a Normal marker and placed at the center of span.
Key Path
Marker ->
Remote Command
:CALCulate:MARKer[1]|2|...12[:SET]:DELTa:CENTer
Example
CALC:MARK2:CENT sets the CF of the analyzer to the value of marker 2.
Notes
Sending this command selects the subopcoded marker
Dependencies
This function is only available when the selected marker is a delta marker. Otherwise the key is
grayed out.
In addition, this function is not available when x-axis is the time domain
Initial S/W Revision
Prior to A.02.00
Mkr ->Span
Sets the start and stop frequencies to the values of the delta markers. That is, it moves the lower
of the two marker frequencies to the start frequency and the higher of the two marker frequencies
to the stop frequency. The marker mode is unchanged and the two markers (delta and reference)
end up on opposite edges of the display.
Key Path
Marker ->
Remote Command
:CALCulate:MARKer[1]|2|...12[:SET]:DELTa:SPAN
Example
CALC:MARK2:DELT:SPAN sets the start and stop frequencies to the values of marker 2 and its
reference marker.
782
Remote Language Compatibility Measurement Application Reference