Agilent Technologies Signal Analyzer N9030a User Manual
Page 148
3 Programming the Analyzer
STATus Subsystem
Example
STAT:QUES:CAL?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Negative Transition
This command determines which bits in the Questionable Calibration Condition register will set
the corresponding bit in the Questionable Calibration Event register when the condition register
bit has a negative transition (1 to 0). The variable
the bits that you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:NTRansition
:STATus:QUEStionable:CALibration:NTRansition?
Example
STAT:QUES:CAL:NTR 16384
Alignment is not required.
Preset
0
Min
0
Max
32767
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Positive Transition
This command determines which bits in the Questionable Calibration Condition register will set
the corresponding bit in the Questionable Calibration Event register when the condition register
bit has a positive transition (0 to 1). The variable
the bits that you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:PTRansition
:STATus:QUEStionable:CALibration:PTRansition?
Example
STAT:QUES:CAL:PTR 16384
Alignment is required.
148