Scale / div, Scale type – Agilent Technologies Signal Analyzer N9030a User Manual
Page 529
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale
Scale / Div
Sets the units per vertical graticule division on the display. This function is only available when
Scale Type (Log) is selected and the vertical scale is power. When Scale Type (Lin) is selected,
Scale/Div is grayed out.
Key Path
AMPTD Y Scale
Remote Command
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:PDIVision
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:PDIVision?
Example
DISP:WIND:TRAC:Y:PDIV 5 DB
Dependencies
Scale/Div is grayed out in linear Y scale. Sending the equivalent SCPI command does change the
Scale/Div, though it has no affect while in Lin.
Preset
10.00 dB / Div
State Saved
Saved in instrument state
Min
0.10 dB
Max
20 dB
Initial S/W Revision
Prior to A.02.00
Scale Type
Chooses a linear or logarithmic vertical scale for the display and for remote data readout.
When Scale Type (Log) is selected, the vertical graticule divisions are scaled in logarithmic units.
The top line of the graticule is the Reference Level and uses the scaling per division Scale/Div to
assign values to the other locations on the graticule.
When Scale Type (Lin) is selected, the vertical graticule divisions are linearly scaled with the
reference level value at the top of the display and zero volts at the bottom. Each vertical division
of the graticule represents one-tenth of the Reference Level.
The Y Axis Unit used for each type of display is set by pressing Y Axis Unit. The analyzer remembers
separate Y Axis Unit settings for both Log and Lin.
Key Path
AMPTD Y Scale
Remote Command
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:SPACing LINear | LOGarithmic
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:SPACing?
Example
DISP:WIND:TRAC:Y:SPAC LOG
DISP:WIND:TRAC:Y:SPAC?
Remote Language Compatibility Measurement Application Reference
529