Mkpx (marker peak excursion), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 401: Description, Mkpx (marker peak
5 Legacy Command Descriptions
MKPX (Marker Peak Excursion)
MKPX (Marker Peak Excursion)
Syntax
Preset State: 6 dB
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
Specifies the minimum signal excursion for the instrument’s internal peak identification routine.
The default value is 6 dB. In this case, any signal with an excursion of less than 6 dB on either
side of the marker would not be identified. Thus, if an MKPK NH command were to be executed on
such a signal, the instrument would not place a marker on this signal peak.
Format
MKPX
MKPX UP|DN
UP or DN increments by one vertical display division
MKPX OA
MKPX?
Query Data Type
SCPI Equivalent Commands
:CALCulate:MARKer:PEAK:EXCursion
)
Remote Language Compatibility Measurement Application Reference
401