6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector
State Saved
The type of each trace is saved in instrument state
Initial S/W Revision
Prior to A.02.00
Remote Command
:TRACe[1]|2|...6:MODE WRITe|MAXHold|MINHold|VIEW|BLANk
:TRACe[1]|2|...6:MODE?
Notes
The legacy TRACe:MODE command is retained for backwards compatibility. In conjunction with
the legacy :AVErage command, it works as follows:
•
:
AVErage ON|OFF sets/clears a variable which we will call average for the sake of this
discussion. This variable is maintained by the analyzer solely for backwards compatibility.
See the [:SENSe]:AVERage[:STATe] command description below.
•
:TRACe:MODEWRITe
sets :TRACe:TYPE WRITe (Clear Write) unless
average is true, in which case it sets it to :TRACe:TYPE AVErage. It also
sets :TRACe:UPDate ON, :TRACe:DISPlay ON, for the selected
trace.
•
:TRACe:MODE MAXHold
sets:TRACe:TYPE MAXHold (Max Hold). It
also sets:TRACe:UPDate ON, :TRACe:DISPlay ON, for the
selected trace.
•
:TRACe:MODE MINHold
sets :TRACe:TYPE MINHold (Min Hold). It
also sets:TRACe:UPDate ON, :TRACe:DISPlay ON, for the
selected trace.
•
:TRACe:MODE VIEW
sets :TRACe:UPDate OFF, :TRACe:DISPlay ON, for
the selected trace
•
:TRACe:MODE BLANk
sets :TRACe:UPDate OFF, :TRACe:DISPlay OFF,
for the selected trace
The query will return the same value as a :TRACe:TYPE? Query, meaning that if you set
:TRACe:MODE:VIEW or :TRACe:MODE:BLANk, the query response will
not be what you sent.
Preset
WRITe
State Saved
The trace mode is an alias only
Backwards Compatibility
Notes
The legacy command :TRACe[n]:MODE was formerly used to set the type or “writing mode” of
the trace. At that time, View and Blank were writing modes. The new TRACe:TYPE command
should be used in the future, but TRACe:MODE is retained to afford backwards compatibility.
Initial S/W Revision
Prior to A.02.00
Remote Command
[:SENSe]:AVERage[:STATe] ON|OFF|1|0
[:SENSe]:AVERage[:STATe]?
Preset
OFF
1322
Remote Language Compatibility Measurement Application Reference