Agilent Technologies Signal Analyzer N9030a User Manual
Page 785
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas
The INITiate command works in a slightly different fashion in the Spectrogram view. In the other Views
(Normal, Trace Zoom and Zone Span), the following two commands perform exactly the same function:
:INITiate:RESTart
:INITiate:IMMediate
However, in the Spectrogram View, the command :INITiate:RESTart works like the Restart key, and
clears out the Spectrogram trace history. The command :INITiate:IMMediate does not clear out the
Spectrogram trace history but performs all other functions of performing a restart.
The table below lists the measurement commands and their responses for the SANalyzer
measurement. Note that the marker values are x, y pairs.
Command
n
Return Value
INITiate:SANalyzer
n/a
n/a
CONFigure?
n/a
long form name of current measurement, for example,
"SANalyzer"
CONFigure:SANalyzer
n/a
n/a (selects SAN measurement with Meas Setup settings
in preset state – same as Meas Preset)
CONFigure:SANalyzer:NDEFault
n/a
n/a (selects SAN measurement without affecting settings)
FETCh:SANalyzer[n]?
MEASure:SANalyzer[n]?
READ:SANalyzer[n]?
0
Returns the following comma-separated results:
1. 1 if there is any margin or limit failure, otherwise 0
2. 0 (future).
3. 0 (future).
4. 0 (future).
5. N dB points result (not a number if off)
6. Current average count k (the current number of data
measurements that have already been combined, in the
averaging calculation).
7. Number of points in the sweep
8. 0 (future).
9. 0 (future).
10. 0 (future).
11. Marker 1 value (x,y)
12. Marker 2 value (x,y)
Remote Language Compatibility Measurement Application Reference
785