Free run – Agilent Technologies Signal Analyzer N9030a User Manual
Page 1400
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
The following examples show trigger setup parameters using an external trigger source.
Example 1 illustrates the trigger conditions with negative slope and no trigger occurs during
trigger Holdoff time.
Example 2 illustrates the trigger conditions with positive slope, trigger delay, and auto trigger
time.
Free Run
Pressing this key, when it is not selected, selects free-run triggering. Free run triggering occurs
immediately after the sweep/measurement is initiated.
Key Path
Trigger
Example
TRIG:SOUR IMM Swept SA measurement
TRIG:
State Saved
Saved in instrument state
Status Bits/OPC
dependencies
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
1400
Remote Language Compatibility Measurement Application Reference