6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
This is unaffected by Preset but is set to the default value on a "Restore Input/Output Defaults"
or "Restore System Defaults->All"
Range
50 Ω | 1 MΩ
Initial S/W Revision
Prior to A.02.00
Q Skew
Sets the skew factor for the Q channel. The skew will shift the channel's data in time. Use this to
compensate for differences in the electrical lengths of the input paths due to cabling and probes.
Key Path
Input/Output, I/Q, Q Setup
Remote Command
[:SENSe]:CORRection:IQ:Q:SKEW
[:SENSe]:CORRection:IQ:Q:SKEW?
Example
Delay the data for the Q channel by 10 ns.
CORR:IQ:Q:SKEW 10 ns
Preset
0
State Saved
Yes
This is unaffected by a Preset but is set to the default value on a "Restore Input/Output
Defaults" or "Restore System Defaults->All"
Range
0 s to 100 ns
Min
0 s
Max
+100 ns
Initial S/W Revision
Prior to A.02.00
Q Probe
Accesses the probe setup parameters for the Q channel. See
Key Path
Input/Output, I/Q, Q Setup
State Saved
No
Readback Text
[]
This is reporting the type of probe sensed on the Q port. There is no parameter for overriding
what is sensed.
Initial S/W Revision
Prior to A.02.00
606
Remote Language Compatibility Measurement Application Reference