6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Source

Source

There is no Source control functionality for this measurement. When this key is pressed, the
screen either displays a blank menu, or the previously-selected menu remains unchanged.

Key Path

Front-panel key

RF Output

Allows you to turn the source RF Power on or off.

As stated below, when the RF Output is turned on, the Source Mode is set to Tracking.  See the

"Source

Mode" on page 1133

key description for special considerations concerning how to configure your

N5172B or N5182B source for use with External Source Control.

Key Path

Source

Scope

Meas Global

Remote Command

:OUTPut[:EXTernal] [:STATe] ON|OFF|1|0

:OUTPut[:EXTernal] [:STATe]?

Example

:OUTP ON

:OUTP?

Dependencies

Grayed out in measurements that do not support a source.  If you go to such a measurement the
output will be forced to Off. 

Grayed out if there is no valid source selection, in this case go to the Select Source menu  to
choose, configure and/or verify your source

When there is no available Source Mode (other than Off), due to other couplings, then the RF
Ouput key is grayed out.

Couplings

When RF Output is turned On, Source Mode is set to Tracking

When Source Mode is turned Off, RF Output is turned Off.

When Source Mode is turned Off (or forced to Off by another coupling), RF Output is turned Off.

Turning RF Output Off does not affect Source Mode or other settings.

Preset

OFF   (on either a Mode Preset, a Source Preset, or Restore Input/Output Defaults)

State Saved

Part of the Input/Output system, which means it is Loaded and Saved with state.

Range

On | Off

Initial S/W Revision

A.06.01

Modified at S/W
Revision

A.10.01