6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker
•
If a Marker Function is on, the X Axis value and the Y-axis function result the marker had
when it became fixed.
See
"Fixed Marker X Axis Value" on page 679
See
"Fixed Marker Y Axis Value" on page 679
See
Fixed Marker X Axis Value
Key Path
Marker, Fixed
Example
:CALC:MARK:MODE FIX sets Marker 1 to Fixed.
Notes
See the description under the “ Marker” key, above.
Dependencies
•
You cannot directly set the X or Y value of a Fixed marker which has a marker function turned
on. If an attempt is made to actually adjust it while a Marker Function is on, a warning
message is generated.
•
You cannot directly set the Y value of a Fixed marker while Normalize is turned on. If an
attempt is made to do so while Normalize is on, a warning message is generated.
State Saved
The marker control mode (Normal, Delta, Fixed, Off) and X and Y Axis values are saved in
instrument state
Backwards Compatibility
Notes
In legacy analyzers, only a Reference marker could be Fixed, and it was always Fixed.
Additionally it could not be noved. In the X-Series, any marker can be set to Fixed and can be
moved to any X or Y value.
Initial S/W Revision
Prior to A.02.00
Fixed Marker Y Axis Value
Key Path
Marker, Fixed
Example
:CALC:MARK:MODE FIX sets Marker 1 to Fixed.
Notes
See the description under the Marker key.
Dependencies
You cannot directly set the X or Y value of a Fixed marker which has a marker function turned
on. If an attempt is made to actually adjust it while a Marker Function is on, a warning message
is generated.
State Saved
The marker control mode (Normal, Delta, Fixed, Off) and X and Y Axis values are saved in
instrument state
Default Unit
depends on the current selected Y axis unit
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference