Agilent Technologies Signal Analyzer N9030a User Manual
Page 1410
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
Key Path
Trigger
Example
TRIG:SOUR EXT2 Swept SA measurement
TRIG:
Dependencies
In some models, there is no second External input. In these models, the External 2 key is
blanked and the EXTernal2 parameter will generate a “Hardware missing; Not available for this
model number” message.
Grayed out if in use by Point Trigger in the Source Setup menu.
Forced to Free Run if already selected and Point Trigger is set to External 2.
State Saved
Saved in instrument state
Status Bits/OPC
dependencies
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.03.00
Trigger Level
Sets the value where the external 2 trigger input will trigger a new sweep/measurement.
Key Path
Trigger, External 2
Remote Command
:TRIGger[:SEQuence]:EXTernal2:LEVel
:TRIGger[:SEQuence]:EXTernal2:LEVel?
Example
TRIG:EXT2:LEV 1.1 V
Couplings
This same level is used for the Ext2 trigger source in the Trigger menu, for the Ext2 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext2 selection in the Gate Source menu.
Preset
1.2 V
State Saved
Saved in instrument state
Min
–5 V
Max
5 V
Default Unit
V
Backwards
:TRIGger[:SEQuence]:FRAMe:EXTernal2:LEVel
1410
Remote Language Compatibility Measurement Application Reference