Agilent Technologies Signal Analyzer N9030a User Manual
Page 1130
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Source
Source Frequency = (Analyzer Frequency *Multiplier Numerator / Multiplier Denominator) +
Source Frequency Offset
Analyzer Frequency is the frequency to which the analyzer is set, which is the analyzer’s
displayed frequency, offset by any Freq Offset set under the Frequency hardkey. Source Frequency
Offset is the value set under Source, Frequency, Freq Offset.
Key Path
Source
Readback
none in Tracking Source mode
Initial S/W Revision
A.06.01
Multiplier Numerator
The multiplier numerator parameter offsets the source frequency from the analyzer frequency.
The source frequency tracks the SA frequency according to the source frequency equation shown
under the Source, Frequency key description.
The multiplier numerator must be restricted to operate within the range of the source minimum
and maximum frequencies.
Key Path
Source, Frequency
Mode
SA
Scope
Mode Global
Remote Command
:SOURce:FREQuency[:MULTiplier]:NUMerator
:SOURce:FREQuency[:MULTiplier]:NUMerator?
Example
:SOUR:FREQ:NUM 3
Sets the source frequency multiplier numerator to 3.
Dependencies
If the currently selected source does not support this capability (for example, an internal
Tracking Generator which must track the LO), this key is forced to its Preset value and grayed
out
Preset
This is unaffected by Mode Preset but is set to 1 on a Source Preset or Restore Input/Output
Defaults.
State Saved
Part of the Input/Output system, which means it is Loaded and Saved with state.
Min
1
Max
1000
Initial S/W Revision
A.06.01
1130