6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
•
If Holdoff On and Type = Normal, readback value
•
If Holdoff On and Type = Above, readback value followed by AL
•
If Holdoff On and Type = Below, readback value followed by BL
•
If Holdoff Type selection is not supported by the current measurement, Holdoff Type is always
Normal
Initial S/W Revision
A.02.00
Auto Trig
Sets the time that the analyzer will wait for the trigger conditions to be met. If they are not met
after that much time, then the analyzer is triggered anyway.
Key Path
Trigger, Auto/Holdoff
Remote Command
:TRIGger[:SEQuence]:ATRigger
:TRIGger[:SEQuence]:ATRigger?
:TRIGger[:SEQuence]:ATRigger:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:ATRigger:STATe?
Example
TRIG:ATR:STAT ON
TRIG:ATR 100 ms
Notes
The "time that the analyzer will wait" starts when the analyzer is ready for a trigger, which may
be hundreds of ms after the data acquisition for a sweep is done. The "time" ends when the
trigger condition is satisfied, not when the delay ends.
Preset
Off, 100 ms
State Saved
Saved in instrument state
Min
1 ms
Max
100 s
Default Unit
s
Initial S/W Revision
Prior to A.02.00
Trig Holdoff
Sets the holdoff time between triggers. When the trigger condition is satisfied, the trigger occurs,
the delay begins, and the holdoff time begins. New trigger conditions will be ignored until the
holdoff time expires. For a free-running trigger, the holdoff value is the minimum time between
triggers.
Remote Language Compatibility Measurement Application Reference