Clear all traces, Preset all traces, Math – Agilent Technologies Signal Analyzer N9030a User Manual
Page 1360
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector
Key Path
Trace/Detector
Remote Command
:TRACe:CLEar TRACE1|TRACE2|TRACE3|TRACE4|TRACE5|TRACE6
Example
TRAC:CLE TRACE1 Clears trace 1
Initial S/W Revision
Prior to A.02.00
Clear All Traces
Clears all traces. Does not affect the state of any function or variable in the instrument. Loads
mintracevalue into all of the points all traces, except traces in Min Hold in which case it loads
maxtracevalue. Does so even if Update=Off.
Key Path
Trace/Detector
Remote Command
:TRACe:CLEar:ALL
:TRACe:PRESet:ALL
When Signal ID is on, this key is grayed out.
Example
TRAC:CLE:ALL Clears all traces
Initial S/W Revision
Prior to A.02.00
Preset All Traces
Turns on Trace 1 and blanks all other traces. Useful when you have many traces on and you
want to go back to having only Trace 1 on the display. Does not affect the trace type, detector or
any other aspect of the trace system.
Key Path
Trace/Detector
Remote Command
:TRACe:PRESet:ALL
When Signal ID is on, this key is grayed out.
Example
TRAC:PRES:ALL
Initial S/W Revision
A.08.01
Math
This menu lets you turn on trace math functions. Trace math functions perform mathematical
operations between traces and, in some cases, user-specified offsets. When in a trace math
1360
Remote Language Compatibility Measurement Application Reference