6 RLC Swept SA Measurement Front-Panel & SCPI Reference
AMPTD Y Scale
trace 1 is on the fifth graticule line from the top, it represents 5 dBm and will read out remotely
as 5.
Example 2, set the following:
Scale Type (Lin)
Y Axis Unit, Volts
Ref Level, 100 mV (10 mV/div)
This sets the top line to 100 mV and the bottom line to 0 V, so each vertical division represents 10
mV. Thus, if a point on trace 1 is on the fifth graticule line from the top, it represents 50 mV and
will read out remotely as 50.
Dependencies
If an amplitude correction with an Antenna Unit other than None is applied and enabled, then
that antenna unit is forced and the key with that unit is the only Y Axis Unit available. All other Y
Axis Unit keys are grayed out.
If an amplitude correction with an Antenna Unit other than None is applied and enabled, and you
then turn off that correction or set Apply Corrections to No, the Y Axis Unit that existed before
the Antenna Unit was applied is restored.
Couplings
The analyzer retains the entered Y Axis Unit separately for both Log and Lin amplitude scale
types
Preset
dBm for log scale, V for linear. The true ‘preset’ value is dBm, since at preset the Y Scale type is
set to logarithmic.
State Saved
Saved in instrument state
Readback line
1-of-N selection
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.02.00, A.04.00, A.11.00
dBm
Sets the amplitude unit for the selected amplitude scale (log/lin) to dBm.
Key Path
AMPTD Y Scale, Y Axis Unit
Example
UNIT:POW DBM
Dependencies
Grayed out if an Amplitude Correction with an Antenna Unit is ON.
Readback
dBm
Initial S/W Revision
Prior to A.02.00
Modified at S/W
Revision
A.02.00