beautypg.com

Acpalpha (adjacent channel power alpha weighting), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual

Page 213: Description, Syntax legacy products

background image

5  Legacy Command Descriptions

ACPALPHA (Adjacent Channel Power Alpha Weighting)

ACPALPHA (Adjacent Channel Power Alpha Weighting)

Syntax

Legacy Products

8560 series

Description

Sets the alpha weighting for ACP measurements.

Format

ACPALPHA (Valid Range from 0 to 1)

ACPALPHA?

Query Data Type

(Valid Range from 0 to 1)

SCPI Equivalent Commands

None

Preset

Default – 0.35

Not affected by preset or Power Cycle

Couplings

Errors

Notes

The functions of the command A2 are identical to the MXMH TRA command. See

"MXMH

(Maximum Hold)" on page 415

.

Remote Language Compatibility Measurement Application Reference

213