6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Trigger
External 1
Pressing this key, when it is not selected, selects the external input port that you will use for the
periodic trigger synchronization. Pressing this key, when it is already selected, accesses the
external 1 sync source setup menu.
Key Path
Trigger, Periodic Timer, Sync Source
Example
TRIG:FRAM:SYNC EXT
Couplings
Same as External 1 trigger source.
Readback
External 1
Initial S/W Revision
Prior to A.02.00
Trigger Level
Sets the value where the external 1 trigger input will trigger a new sweep/measurement.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:LEVel
:TRIGger[:SEQuence]:EXTernal1:LEVel?
Example
TRIG:EXT1:LEV 0.4 V
Couplings
This same level is used for the Ext1 trigger source in the Trigger menu, for the Ext1 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext1 selection in the Gate Source menu.
Preset
1.2 V
State Saved
Saved in instrument state
Min
–5 V
Max
5 V
Default Unit
V
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:EXTernal:LEVel
For backward compatibility, the parameter EXTernal is mapped to EXTernal1
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal1:LEVel
Initial S/W Revision
Prior to A.02.00