6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
2. When INPut:MIXer INTernal is received, SENSe:FEED RF is executed.
3. When INPut:MIXer? is received, the response will be INT if any input other than the external
mixer is selected and EXT if the external mixer is selected
Preset
INT
Backwards Compatibility
Notes
PSA supports the following SCPI Command :
:INPut:MIXer:TYPE PRESelected|UNPReselect
:INPut:MIXer:TYPE?
PXA does not support the :INPut:MIXer:TYPE command.
Initial S/W Revision
A.08.01
Input/Output variables - Preset behavior
Virtually all the input/output settings are not a part of mode preset. They can be set to their
default value by one of the three ways:
l
by using the Restore Input/Output Defaults key on the first page of the input/output menu,
l
by using the System->Restore System Defaults->Input/Output Settings or,
l
by using the System -> Restore System Defaults->All. Also, they survive a Preset and a Power
cycle.
A very few of the Input/Output settings do respond to a Mode Preset; for example, if the
Calibrator is on it turns off on a Preset, and if DC coupling is in effect it switches to AC on a
Preset. These exceptions are made in the interest of reliability and usability, which overrides the
need for absolute consistency. Exceptions are noted in the SCPI table for the excepted functions.
RF Input
Selects the front-panel RF input port to be the analyzer signal input. If RF is already selected,
pressing this key accesses the RF input setup functions.
Key Path
Input/Output
Example
[:SENSe]:FEED RF
Readback
The RF input port, RF coupling, and current input impedance settings appear on this key as:
"XX, YY, ZZ" where
XX is RF, RF2, RFIO1, RFIO2, depending on what input is selected (only appears on analyzers with
multiple RF inputs)
Remote Language Compatibility Measurement Application Reference