6 RLC Swept SA Measurement Front-Panel & SCPI Reference
BW
Key Path
BW, RBW Control, Filter BW
Example
BAND:TYPE DB3
Readback
–3 dB
Initial S/W Revision
Prior to A.02.00
–6 dB
Selects the filter bandwidths where the bandwidth is defined at the –6 dB points. This uses the
normal RBW filters, but the value displayed on the key, active function line and screen
annotation changes to reflect the –6 dB bandwidth instead of the –3 dB bandwidth.
Key Path
BW, RBW Control, Filter BW
Example
BAND:TYPE DB6
Readback
–6 dB
Initial S/W Revision
Prior to A.02.00
Noise
Selects the noise filter bandwidths. This uses the normal RBW filters, but the value displayed on
the key, active function line and screen annotation changes to reflect the equivalent noise
bandwidth, instead of the –3 dB bandwidth.
Key Path
BW, RBW Control, Filter BW
Example
BAND:TYPE NOIS
Readback
Noise
Initial S/W Revision
Prior to A.02.00
Impulse
Selects the impulse bandwidths. This uses the normal RBW filters, but the value displayed on the
key, active function line and screen annotation changes to reflect the equivalent impulse
bandwidth instead of the –3 dB bandwidth.
Key Path
BW, RBW Control, Filter BW
Example
BAND:TYPE IMP
Readback
Impulse
Initial S/W Revision
Prior to A.02.00