6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Meas Setup
Demod State (Remote Command Only)
Sets or queries the state of the Analog Demod Tune and Listen function. Setting the state to ON
with this command will select AM demodulation by default and activate it (turn it on).
The response to the query is determined by the current setting of [:SENSE]:DEMod
AM|FM|PM|OFF. The response will be 1 if AM, FM, PM are selected, or 0 if OFF is selected..
Remote Command
[:SENSe]:DEMod:STATe OFF|ON|0|1
[:SENSe]:DEMod:STATe?
Preset
OFF
Initial S/W Revision
Prior to A.02.00
Noise Source
This menu allows you to turn the noise source power on or off when making manual noise figure
measurements.
See
"More Information" on page 854
Key Path
Meas Setup
Remote Command
:SOURce:NOISe:TYPE NORMal | SNS
:SOURce:NOISe:TYPE?
Example
SOUR:NOIS:TYPE NORM
Couplings
If no SNS is connected, this parameter will be set to “Normal”
When Type is set to “SNS” and the SNS is disconnected, this parameter gets bumped to
“Normal”
When an SNS is not connected, the SNS type will be grayed (disabled).
Preset
Normal
State Saved
Saved in instrument state.
Range
Normal | SNS
Backwards Compatibility
Notes
In previous Noise Figure analysis applications, this command
could optionally be preceded with the :SENSe keyword. The
optional :SENSe keyword is no longer supported.
Initial S/W Revision
Prior to A.02.00
Remote Language Compatibility Measurement Application Reference