Agilent Technologies Signal Analyzer N9030a User Manual
Page 151
3 Programming the Analyzer
STATus Subsystem
:STATus:QUEStionable:CALibration:SKIPped:NTRansition?
Example
STAT:QUES:CAL:SKIP:NTR 1
Align RF skipped is not required.
Preset
0
Min
0
Max
32767
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Skipped Positive Transition
This command determines which bits in the Questionable Calibration Skipped Condition register
will set the corresponding bit in the Questionable Calibration Skipped Event register when the
condition register bit has a positive transition (0 to 1). The variable
decimal values of the bits that you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:SKIPped:PTRansition
:STATus:QUEStionable:CALibration:SKIPped:PTRansition?
Example
STAT:QUES:CAL:SKIP:PTR 1
Align RF skipped is required.
Preset
32767
Min
0
Max
32767
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Extended Failure Register
The following commands and queries are available for this register:
"Questionable Calibration Extended Failure Condition" on page 152
"Questionable Calibration Extended Failure Enable" on page 152
"Questionable Calibration Extended Failure Event Query" on page 153
"Questionable Calibration Extended Failure Negative Transition" on page 153
Remote Language Compatibility Measurement Application Reference
151