Agilent Technologies Signal Analyzer N9030a User Manual
Page 633
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Input/Output
Example
:CORR:CSET:ANT GAUS
Readback
" dBG"
Initial S/W Revision
A.02.00
dBµA
Sets the antenna unit to dBµA. If this correction is turned on, and Apply Corrections is on, the Y
Axis Unit will then be forced to dBµA and all other Y Axis Unit selections will be grayed out.
Key Path
Input/Output, Corrections, Properties, Antenna Unit
Example
:CORR:CSET:ANT UA
Readback
" dBµA"
Initial S/W Revision
A.11.00
Frequency Interpolation
This setting controls how the correction values per-bucket are calculated. We interpolate between
frequencies in either the logarithmic or linear scale.
This setting is handled and stored individually per correction set.
See
Key Path
Input/Output, Corrections, Properties
Remote Command
[:SENSe]:CORRection:CSET[1]|2|...6:X:SPACing LINear | LOGarithmic
[:SENSe]:CORRection:CSET[1]|2|...6:X:SPACing?
Example
CORR:CSET:X:SPAC LIN
Preset
Unaffected by a Preset. Set to Linear by Restore Input/Output Defaults.
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00
Interpolation
For each bucket processed by the application, all of the correction factors at the frequency of
interest (center frequency of each bucket) are summed and added to the amplitude. All trace
operations and post processing treat this post-summation value as the true signal to use.
Remote Language Compatibility Measurement Application Reference
633