Ksz (reference level offset), Syntax, Legacy products – Agilent Technologies Signal Analyzer N9030a User Manual
Page 351: Description, Syntax legacy products
5 Legacy Command Descriptions
KSZ (Reference Level Offset)
KSZ (Reference Level Offset)
Syntax
Legacy Products
8566A/B, 8568A/B
Description
Offsets all amplitude readouts on the display but without affecting the trace.
Once activated, KSZ displays the amplitude offset on the left side of the screen.
Sending KSZ 0, or presetting the instrument, eliminates an amplitude offset.
Format
KSZ
8566A/B only supports unit DM
KSZ OA
KSZ?
Query Data Type
SCPI Equivalent Commands
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet?
(See
Preset
0
Notes
The functions of KSZ are identical to
"ROFFSET (Reference Level Offset)" on page 452
Remote Language Compatibility Measurement Application Reference
351