Agilent Technologies Signal Analyzer N9030a User Manual
Page 147
3 Programming the Analyzer
STATus Subsystem
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:CONDition?
Example
STAT:QUES:CAL:COND?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Enable
This command determines which bits in the Questionable Calibration Condition Register will set
bits in the Questionable Calibration Event register, which also sets the Calibration Summary bit
(bit 8) in the Questionable Register. The variable
bits you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:ENABle
:STATus:QUEStionable:CALibration:ENABle?
Example
STAT:QUES:CAL:ENAB 16384 Can be used to query if an alignment is needed, if you have turned
off the automatic alignment process.
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Event Query
This query returns the decimal value of the sum of the bits in the Questionable Calibration Event
register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration[:EVENt]?
Remote Language Compatibility Measurement Application Reference
147