Agilent Technologies Signal Analyzer N9030a User Manual
Page 1220
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Sweep/Control
The Status Operation Register bit 5 "Waiting for Trigger" is set at the same time as the Sweeping
or Measuring bit is set. It is cleared when the trigger actually occurs (that is, after the trigger
event occurs and all the applicable trigger criteria have been met). A corresponding pop-up
message ("Waiting for trigger") is generated if no trigger signal appears after approximately 2
sec. This message goes away when a trigger signal appears.
Initial S/W Revision
Prior to A.02.00
Trigger Level
Sets the value where the external 1 trigger input will trigger a new sweep/measurement.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:LEVel
:TRIGger[:SEQuence]:EXTernal1:LEVel?
Example
TRIG:EXT1:LEV 0.4 V
Couplings
This same level is used for the Ext1 trigger source in the Trigger menu, for the Ext1 selection in
the Periodic Timer sync source (in the Trigger menu and in the Gate Source menu), and also for
the Ext1 selection in the Gate Source menu.
Preset
1.2 V
State Saved
Saved in instrument state
Min
–5 V
Max
5 V
Default Unit
V
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:EXTernal:LEVel
For backward compatibility, the parameter EXTernal is mapped to EXTernal1
Backwards
Compatibility SCPI
:TRIGger[:SEQuence]:FRAMe:EXTernal1:LEVel
Initial S/W Revision
Prior to A.02.00
Trig Slope
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on
a falling edge.
Key Path
Trigger, External 1
Remote Command
:TRIGger[:SEQuence]:EXTernal1:SLOPe POSitive|NEGative
1220
Remote Language Compatibility Measurement Application Reference