6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Mode Preset

does a User Preset.  The old PRESet:TYPE command is ignored (without generating an error), and
SYST:PRES without a parameter does a Mode Preset, which should cover most backward code
compatibility issues.

The settings and correction data under the Input/Output front-panel key (examples: Input Z Corr,
Ext Amp Gain, etc.) are no longer part of any Mode, so they will not be preset by a Mode Preset.
They are preset using Restore Input/Output Defaults, Restore System Defaults All.  Note that
because User Preset does a Recall State, and all of these settings are saved in State, they ARE
recalled when using User Preset.

Initial S/W Revision

Prior to A.02.00

How-To Preset

The table below shows all possible presets, their corresponding SCPI commands and front-panel
access (key paths). Instrument settings depend on the current measurement context. Some settings
are local to the current measurement, some are global (common) across all the measurements in
the current mode, and some are global to all the available modes. In a similar way, restoring the
settings to their preset state can be done within the different contexts.

l

Auto Couple - is a  measurement local key. It sets all Auto/Man parameter couplings in the
measurement to Auto. Any Auto/Man selection that is local to other measurements in the mode
will not be affected.

l

Meas Preset - is a measurement local key.  Meas Preset resets all the variables local to the
current measurement except the persistent ones.

l

Mode Preset - resets all the current mode's  measurement local and  measurement global
variables except the persistent ones.

l

Restore Mode Defaults - resets ALL the Mode variables (and all the Meas global and Meas local
variables), including the persistent ones.

Type Of Preset

SCPI Command

Front Panel Access

Auto Couple

:COUPle ALL

Auto Couple front-panel key

Meas Preset

:CONFigure:

Meas Setup Menu

Mode Preset

:SYSTem:PRESet

Mode Preset (green key)

Restore Mode Defaults

:INSTrument:DEFault

Mode Setup Menu

Restore All Mode Defaults

:SYSTem:DEFault MODes

System Menu; Restore
System Default Menu

*RST

*RST

not possible (Mode Preset
with Single)