Agilent Technologies Signal Analyzer N9030a User Manual
Page 427
5 Legacy Command Descriptions
OL (Output Learn String)
Byte
Support Information
20
Supported: Trigger Mode, Sweep Mode, TRB Clearwrite status
Unsupported: Recorder Output
21
Supported
22
Unsupported: Fixed decimal 0
23 to 25
Supported
26
Supported: Scale Type, Log Scale Factor, and Display State
Unsupported: XY Recorder
27
Supported
28
Unsupported: Fixed decimal 0
29 to 30
Supported
31
Unsupported: Fixed decimal 0
32 to 37
Supported
38
Unsupported: Fixed decimal 0
39 to 45
Supported
46 to 47
VAVG count limit, value returned is always current count value * 2.
48 to 53
Supported
54 to 57
If active marker is a delta marker, active marker absolute Y position only supported for X-Series
58 to 61
If active marker is a delta marker, reference marker absolute Y position only supported for X-Series
62
Unsupported: Fixed decimal 0
63
Supported
64
Supported: Log Amp Units, R3, R2, and R4
Unsupported: Stop sweep
65
Supported: Lin Amp Units, TRC View Status
Unsupported: Bits 5 and 4 (always set HI)
66 to 71
Supported
72
Unsupported: Fixed decimal 0
73
Supported: Video Avg
Unsupported: Power on last, Ext Ref Lvl, Fast HP-IB, Bit 4 (always set HI)
74 to 77
Unsupported: Fixed decimal 0
78
Unsupported
79
Unsupported: Fixed decimal 0
80
Fixed decimal 162
Remote Language Compatibility Measurement Application Reference
427