Marker to, Mkr->cf – Agilent Technologies Signal Analyzer N9030a User Manual
Page 778
6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Marker To
Marker To
The Marker -> key accesses menu keys that can copy the current marker value into other
instrument parameters (for example, Center Freq). The currently selected marker is made the
active function on entry to this menu (if the currently selected marker is not on when you press
this front panel key, it will be turned on at the center of the screen as a normal type marker and
then made the active function).
The Marker -> feature is used to quickly assign a marker’s x- or y-axis value to another
parameter. For example, if a marker’s x-axis value is 500 MHz and y-axis value is –20 dBm,
pressing Mkr -> CF assigns 500 MHz to Center Freq and pressing Mkr - >Ref Lvl assigns –20 dBm to
Ref Level
.
Key Path
Front-panel key
Notes
All Marker To functions executed from the front panel use the selected marker’s values, while
all Marker To remote commands specify in the command which marker’s value to use.
Consistent with other remote marker commands, sending a Marker To remote command will
never change which marker is selected.
Initial S/W Revision
Prior to A.02.00
Mkr->CF
Sets the center frequency of the analyzer to the frequency of the selected marker. The marker stays
at this frequency, so it moves to the center of the display. In delta marker mode, this function sets
the center frequency to the x-axis value of the delta marker. When the frequency scale is in log
mode, the center frequency is not at the center of the display.
If the currently selected marker is not on when this key is pressed, it will be turned on at the
center of the screen as a normal type marker.
Key Path
Marker ->
Remote Command
:CALCulate:MARKer[1]|2|...12[:SET]:CENTer
Example
CALC:MARK2:CENT sets the CF of the analyzer to the value of marker 2.
Notes
Sending this command selects the subopcoded marker
If specified marker is off, this command will turn it on at the center of the screen as a normal
type marker.
Dependencies
This function is not available (key is grayed out) when x-axis is the time domain
Couplings
All the usual couplings associated with setting Center Frequency apply.
Initial S/W Revision
Prior to A.02.00
778
Remote Language Compatibility Measurement Application Reference