Agilent Technologies Signal Analyzer N9030a User Manual
Page 164
3 Programming the Analyzer
STATus Subsystem
Questionable Integrity Signal Event Query
This query returns the decimal value of the sum of the bits in the Questionable Integrity Signal
Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:SIGNal[:EVENt]?
Example
STAT:QUES:INT:SIGN?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Signal Negative Transition
This command determines which bits in the Questionable Integrity Signal Condition register will
set the corresponding bit in the Questionable Integrity Signal Event register when the condition
register bit has a negative transition (1 to 0). The variable
values of the bits that you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition
:STATus:QUEStionable:INTegrity:SIGNal:NTRansition?
Example
STAT:QUES:INT:SIGN:NTR 4
Burst found will be reported to the Integrity Summary of the Status Questionable register.
Preset
0
Min
0
Max
32767
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Integrity Signal Positive Transition
This command determines which bits in the Questionable Integrity Signal Condition register will
set the corresponding bit in the Questionable Integrity Signal Event register when the condition
164