Agilent Technologies Signal Analyzer N9030a User Manual
Page 172
3 Programming the Analyzer
STATus Subsystem
Mode
All
Remote
Command
:STATus:QUEStionable:TEMPerature:ENABle
:STATus:QUEStionable:TEMPerature:ENABle?
Example
STAT:QUES:TEMP:ENAB 1
Reference Oscillator Oven Cold will be reported to the Temperature Summary of the Status Questionable
register.
Preset
32767
Min
0
Max
32767
Status
Bits/OPC
dependencies
Sequential command
Initial S/W
Revision
Prior to A.02.00
Questionable Temperature Event Query
This query returns the decimal value of the sum of the bits in the Questionable Temperature
Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature[:EVENt]?
Example
STAT:QUES:TEMP?
Preset
0
Status Bits/OPC dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Temperature Negative Transition
This command determines which bits in the Questionable Temperature Condition register will set
the corresponding bit in the Questionable Temperature Event register when the condition register
bit has a negative transition (1 to 0).
The variable
172
Remote Language Compatibility Measurement Application Reference