6 RLC Swept SA Measurement Front-Panel & SCPI Reference
Recall
Notes
auto return
Dependencies
Only Correction 1 may be used to load a Correction that contains an Antenna Unit other than
None
Preset
Not part of Preset, but is reset to Correction 1 by Restore Input/Output Defaults; survives
shutdown.
State Saved
The current Correction number is saved in instrument state
Initial S/W Revision
A.02.00
Amplitude Correction
These keys let you select which Correction to import the data into. Once selected, the key returns
back to the Import Data menu and the selected Correction number is annotated on the key. The
next step is to select the Open key in the Import Data menu.
Antenna corrections are a particular kind of Amplitude Corrections – they are distinguished in
the corrections file by having the Antenna Unit set to a value other than None. Only Correction 1
supports Antenna Units.
Key Path
Recall, Data, Amplitude Correction
Notes
auto return
Dependencies
Only Correction 1 may be used to load a Correction that contains an Antenna Unit other than
None
Preset
Not part of Preset, but is reset to Correction 1 by Restore Input/Output Defaults; survives
shutdown.
State Saved
The current Correction number is saved in instrument state
Initial S/W Revision
A.02.00
Trace
This key selects Trace as the data type to be imported. When pressed a second time, it brings up
the Trace Menu, which lets you select the Trace into which the data will be imported.
The trace file contains “meta” data which describes the state of the analyzer when the trace was
exported (see
"Trace File Contents" on page 1045
). If the meta data in the file does not match the
current SA state, the “invalid data indicator” (*) is displayed.
Key Path
Recall, Data
Remote Command
:MMEMory:LOAD:TRACe:DATA TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 |
Remote Language Compatibility Measurement Application Reference